AOS (Beam Reducing Telescopes)
takafumi.ushiba - 18:35 Wednesday 12 October 2022 (22456) Print this report
IR power budget on TMSY optical table

[Haba, Komori, Ushiba]

Abstract:
We measured IR power at several places on TMSY and measured ADC cnts of PD and QPDs at that time.

Detail:
After putting a green filter (klog22458), we measured IR beam power at several places on TMSY optical table.
Measurement results are as follows (optical layout of TMSY can be seen here):
1. In front of RLNS2 (just after the green filter): 480uW +- 10uW (background: 0.73 +- 0.02 uW)
2. In front of RPD1: 52.5 +- 0.5 uW (background: -0.26 +- 0.01 uW)
3. In front of RQPD1: 116 +- 1 uW (background: 0.15 +- 0.00 uW)
4. In front of RQPD2: 118 +- 1 uW (background: 0.53 +- 0.00 uW)

We also measured PD and QPD cnts at each segments when IR was locked.
Following tables are results.

  IR locked Background
PDA1 272.38 17.89
QPDA1 SEG1 -680.33 -29.98
QPDA1 SEG2 -647.87 -30.09
QPDA1 SEG3 -643.04 -29.07
QPDA1 SEG4 -763.46 -29.89
QPDA2 SEG1 -581.50 -26.75
QPDA2 SEG2 -856.75 -28.08
QPDA2 SEG3 -608.52 -26.27
QPDA2 SEG4 -816.47 -26.94

Then, I set offset for background subtruction of QPD segments according to the measurement results.

Note:
In front of PDA1, there is a mirror for reducing IR power on PD (I don't know how much power was thorown away).
Between RST3and RBS2, there was a mirror for reducing laser power on QPDs but it was removed during the work because IR power on QPD is too small.
After removing the mirror at QPD path, I changed the QPD gain to low gain mode for avoid saturation of QPD signals.