tatsuki.washimi - 16:01 Wednesday 05 October 2022 (22353) Print this report
EX/EY Dsub15 -> MAG13, PTB110

[YokozaWashimi]

We replaced the connector for IY0 Dsub15 -> MAG13, PTB110 from the temporally one to the new one.
We also located the Barometers, VAISALA PTB110 for EX 1F and EY 1F.

All of their signals look healthy.

(Strange signal seen yesterday was due to the whitening filter being applied and 60 Hz noise being enhanced.)

Images attached to this report
Comments to this report:
tatsuki.washimi - 8:55 Thursday 06 October 2022 (22373) Print this report

plot update

Images attached to this comment
tatsuki.washimi - 17:04 Friday 07 October 2022 (22410) Print this report

coherence with infrasound microphones

Images attached to this comment