[Terri, Keiko, Chiaki]
The calculated power entering the TSM table is 677uW.
The power could not be measured after the periscope as the beam was too large.
We measured 538uW after the first lens and 525uW after the second lens on the table, so estimate ~550uW entering the table.
Our calculation is an estimate and there may be additional loss in the arm or on transmission through the mirror so the power seems reasonable and therefore we cannot conclude any losses from birefringence.(Written by Terri)
We also repositioned the optics after lens2 for the optical axis of future QPDs.
(TMSX_20221003.png : blue power->measurement power, blue number: number of opitical tabel hole : ../../users/Commissioning/data/ASC/2022/TMSX/BeamProfile/20221003/KAGRA_TMS_Xarm.sng)
This figure shows the PD signal of TRANS IR before and after placing PD. (TMSX_IR_power_before_after_placing.png)