Aso, Hirata, Nishino
@Locked state
Before SRM: 52(3) mW*
After SRM; 52(3) mW*
After OMMT1; 52(3) mW*
After OMMT2 (= before OFI): 49(2) mW
After OFI: 46(2) mW -> OFI transimissivity = 94(6) %
After OSTM (rellection): 45(1) mW -> OSTM reflectivity = 98(5) %
OSTM (transmission): 0.42(1) mW -> OSTM transmissivity = 0.91(5) %
Total transimissivity from before-OFI to after-OSTM: 92(4) %
@Unlocked state
After OMMT2 (= before OFI): 62(1) mW
After OFI: 58(1) mW -> OFI transimissivity = 94(2) %
After OSTM (rellection): 57(1) mW -> OSTM reflectivity = 98(2) %
OSTM (transmission): 0.52(1) mW -> OSTM transmissivity = 0.91(3) %
Total transimissivity from before-SRM to after-OSTM: 92(5) %
MEMO
*We used two power meters, Ophir Photonics and Thorlabs, and did intercalibration.
Ophir's one has a bigger aperture so is capable of the beam size before around OMMT1 (but power resolusion is worse than Thorlabs).
* indicates the one measured by Ophir Photonics.