VAC (EY)
nobuhiro.kimura - 13:45 Tuesday 31 May 2022 (20956) Print this report
Summaries of the vacuum leak test for EYC on 30~31/May/2022
When: 30~31/May/2022
Who:Kimura, Yasui, M. Takahashi, Sawada (hokuto) and Ueda (SK)
Task:
We found very large vacuum leak place at the duct shield cryo-cooler called Yer at arm side.
The place was on 22 multi pin type hermetic sealed connector .
Pleas see attached photos.
The leak level was 2.0 x 10^-7 Pam^3/s.

We are assuming the reason of leak was crack in the welding part between ICF-70 flange and the connector.
We will discuss how to repair it.
Images attached to this report
Comments to this report:
nobuhiro.kimura - 20:52 Wednesday 01 June 2022 (20981) Print this report
When: 1/June/2022
Who:Kimura, Yasui, M. Takahashi, Sawada (Hokuto) and Ueda (SK)
Task:
We repaired large vacuum leak place level of 2.0 x 10^-7 Pam^3/s by "VACSEAL".
"VACSEAL" was diluted 20-fold with acetone.
A 3 ml volume syringe was used to inject the diluted solution into the leak place.
After the diluted solution was injected, the leak appeared to be sealed.
The amount of leakage was confirmed to be less than the detection sensitivity (10^-13 Pam^3/s)
of the helium leak detector used for the inspection.
The test is scheduled to be conducted again on June 2, when the diluted solution will solidify.
Images attached to this comment