The following measurements were taken to install DC centering control for IMMT1 and IMMT2 using signals received from the POP forward QPDs,
*We measured Sensing matrix of IMMT1, IMMT2 by POP-FORWARD QPDs (20220426_SENSMAT_IMMT{1,2}_PIT.png, 20220426_SENSMAT_IMMT1_PIT.png, 20220426_SENSMAT_IMMT1_2_YAW.png)At this time, I turned off the compensating filter for each QPD since the degrees of freedom are swung at lower frequencies. I also turned off the ADS for IMMT1 and PR2. -->./users/Commisioning/data/ASC/POP_FOR/Sensmat/20220426/TF_POP_FORWARD_QPD_IMMT1{1,2}_{PIT,YAW}.xml
*We inserted the sensing matrix values into DOF3 and DOF4. (Sensmat_POP_FOR.png) and changed the MEDM screen (ASC_CUST_INMTRX_{P, Y}_DC.adl) for them.
*After we checked that the sensors looking at each degree of freedom ( {IMMT1, IMMT2 }_{PIT ,YAW }) were working properly.(Sensor_DOF3_DOF4.png)-->./users/Commisioning/data/ASC/POP_FOR/Sensmat/20220426/Sensmat_POP_FORWARD_QPD_IMMT1{1,2}_{PIT,YAW}.xml
Sensor | DoF |
DoF3 | IMMT1 |
DoF4 | IMMT2 |
*We measured the PLANT transfer function using the sensor we made. (PLANT_POPFOR.png) At this time we turned off DC control for IMMT1 oplev. --> ./users/Commisioning/data/ASC/POP_FOR/PLANT/20220426/PLANT_IMMT{1,2}_{DOF3,4}_{PIT,YAW}.xml