Reports 1-1 of 1 Clear search Modify search
VIS (General)
ryutaro.takahashi - 15:59 Wednesday 13 April 2022 (20471) Print this report
Signal generator for LVDT driver

I checked the status of the signal generators for the LVDT driver. Though the embedded generators (EB) were employed for Type-A, some stand-alone generators are still used. Though the output of the AFG3102G in PRM/PR3 is set to 7.0Vp-p on the interface, the actual amplitude is 12.0Vp-p. This is too large, isn't it?

  LVDT driver SG REF LOOP[Vp-p]
IX GAS/IP/BFD EB 6.2
  ACC FG-274 6.2
IY GAS/IP/BFD EB 6.0
  ACC EB 6.2
EX GAS/IP/BFD DG1022A 6.5
  ACC EB 7.0
EY GAS/IP EB 6.6
  BFD EB 9.4
  ACC EB 7.0
BS GAS/IP DG1022A 4.6
SRM GAS/IP DG1022A 4.5
SR2 GAS/IP FG-274 5.1
SR3 GAS/IP DG1022A 4.5
PRM GAS/BFD AFG3102G 12.0
PR2 GAS/BFD DG1022A 5.3
PR3 GAS/BFD AFG3102G 12.0
Search Help
×