VAC (EX)
Kimura - 9:58 Tuesday 18 January 2022 (19491) Print this report
Result of the vacuum leak test for EXC area between X-arm gate valve and BRT gate valve (Continued)
When: 17/Jan./2022
Who:Kimura, M. Takahashi and Sawada (Hokuto)
Task:
A vacuum leak test was carried out between the X-arm gate valve and the BRT gate valve in the EXC area, again.
It was confirmed that there was very large vacuum leak of ~1 x 10^-9 Pam^3/s around the EXV chamber.
Pleas see attached first and second photos from display on helium leak detector.

Candidate leak locations are as follows;
#1: Helium gas entered through the hole for the push screw on the EXV horizontal flange and reached the leak occurring at the elastomer seal.
#2: Cracks in the ceramic sealing area of the hermetic seal connector.
Pleas see attached third photo.

Of the above two candidate locations, we could not identify which one was the location of the leak.
We plan to conduct another leak test this Friday (21/Jan./2022) to determine the location of the leak.
Images attached to this report
Comments to this report:
nobuhiro.kimura - 14:55 Saturday 22 January 2022 (19536) Print this report
When: 21/Jan./2022
Who:Kimura, Tomaru, Yasui and Yano
Task:
We fixed very large vacuum leak place at the EXV chamber.
The place was on D-sub type hermetic sealed connector .
Pleas see attached photos.
The leak level was 3.0 x 10^-7 Pam^3/s.

We are assuming the reason of leak was crack in the ceramic or pin hole of sealing between pin and the ceramic.
We will discuss how to repair it.
Images attached to this comment
nobuhiro.kimura - 5:11 Wednesday 26 January 2022 (19561) Print this report
When: 25/Jan./2022
Who: Kimura, M. Takahashi, Yasui and Sawada (Hokuto)
Task:
We found very small vacuum leak place at new vacuum pump near EXT chamber.
The place was on connection flange between tees and Ion pump.
The leak level was 5.0 x 10^-11 Pam^3/s.
Pleas see attached photos.

We are assuming the reason of leak was tilting of the Ion pump by its weight.
We asked Mirapro how to repair it.
Images attached to this comment
takashi.uchiyama - 11:04 Friday 11 February 2022 (19758) Print this report
Mirapro, Takahashi, Sawada, Uchiyama

2022/02/10

We repaired the vacuum leakage at the new vacuum pump near EXT chamber.