Reports 1-1 of 1 Clear search Modify search
VAC (IX)
nobuhiro.kimura - 4:45 Thursday 13 January 2022 (19426) Print this report
Result of the vacuum leak test at IXC
When: 12/Jan./2022
Who:Kimura, Yasui, Tomaru (NAOJ), Sawada (Hokuto) and Ueda (SK service)
Task:
We performed the vacuum leak test for IXC area between BS gate valve and X-arm gate valve.
We confirmed that there were no vacuum leaks larger than 1 x 10^-11 Pam^3/s at the IXC area.
Comments to this report:
nobuhiro.kimura - 5:22 Wednesday 26 January 2022 (19565) Print this report
When: 25/Jan./2022
Who: Kimura, M. Takahashi, Yasui and Sawada (Hokuto)
Task:
We tried to find vacuum leak place around D-sub type hermetic connectors of IXV.
Then, we confirmed there were no leaks the connectors like EXV.

We will continue vacuum leak test at IXC.
nobuhiro.kimura - 8:49 Thursday 27 January 2022 (19592) Print this report
When: 26/Jan./2022
Who: Kimura and Yasui
Task:
We tried to find vacuum leak place between IXV and cross-tube.
Then, we confirmed there were no leaks.

We will continue vacuum leak test at IXC.
nobuhiro.kimura - 8:00 Friday 28 January 2022 (19609) Print this report
When: 27/Jan./2022
Who: Kimura, Yasui, M. Takahashi and Sawad (Hokuto)
Task:
We tried to find vacuum leak place in clean booth area of IXC.
Then, we confirmed there were no leaks.

We will continue vacuum leak test left area of IXC.
nobuhiro.kimura - 8:19 Thursday 03 February 2022 (19658) Print this report
When: 2/Feb./2022
Who: Kimura, Yasui, M. Takahashi and Sawad (Hokuto)
Task:
We tried to find vacuum leak place in clean booth area of IXC.
Then, we concluded there were no vacuum leaks at IX any more.
We will start to prepare duct shield cooling.