VIS (IMM)takaaki.yokozawa - 16:59 Tuesday 04 January 2022 (19346)
Print this reportIMMT1 oplev work[Tomaru, Akutsu, Yasui, Yokozawa]
1. Replaced the SLD for IMMT1 oplev to new one. #203245 -> #242769 http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Subgroups/AOS/Oplev_Light_Sources Before replacement, the laser power was about 1300 count at high QPD gain mode. After replacement, the laser power became 10,000 count.
2. Set the IR filter Sorry I forgot to take picture after installation, we set the IR filter and QPD cover . After setting, the laser power became 9,200 count
Comments to this report:
tomotada.akutsu - 20:25 Tuesday 04 January 2022 (19349)
Print this report
About the QPD cover assembly.
The QPD cover body was manufactured by NAOJ Advanced Technology Center with a mechanical 3D printer, plus additional manufacturing in the machine shop for making the screw part.
The optical filter was to stop 532 nm and 1064 nm in >OD6 (in nominal design) from Sigma koki (see the attached photo and JGW-D2213690)