Reports 1-1 of 1 Clear search Modify search
VIS (SR2)
fabian.arellano - 12:10 Friday 25 June 2021 (17282) Print this report
Comment to IP resonance frequencies (17208)

Following the analysis done in SR3 of the asymmetric values of IP-L and IP-T resonance frequencies (klog 17281), I carried out the same analysis in SR2 in which there is no asymmetry or it's jut too hard to distingush.

I measured the transfer functions at the individual LVDT-actuator units H1, H2 and H3 separately. As can be seen in the plots attached, the lowest resonance frequencies in each case are:

  • H1: between 66 and 70 mHz,
  • H2: 66 mHz,
  • H3: between 66 and 70 mHz.
  • Frequency resolution: 5 mHz.
  • Temperature: 23.6°C (O3GK T = 23.5 °C).

The resonance frequencies are either the same or very close together when measuring with the current frequency resolution.

Images attached to this comment
Search Help
×

Warning

×