Reports 1-1 of 1 Clear search Modify search
IOO (General)
takahiro.yamamoto - 21:12 Tuesday 23 March 2021 (16311) Print this report
Calibration of the PZT actuator efficiency
[Kenta, YamaT]

We measured the actuator efficiency of PZT for PMC.

The result is consistent with klog#5379. (~169V/FSR)

We are now measuring the TF of temperature loop. It will takes more 3 hours.
Don't touch PMC until finishing this measurement.

-----

We swept an offset on PZT slow output from -9V to 9V (It is consistent with 30V to 300V on HV amp. output)

and by changing the HV output as 169V, we observed resonant peak moved one FSR. (See Fig1). 
 
 
Images attached to this report
Comments to this report:
takahiro.yamamoto - 2:26 Thursday 25 March 2021 (16324) Print this report
Re-measurement is ongoing. It will be finished around 8 a.m. (maybe).
Don't restart DAQ tonight.
takahiro.yamamoto - 3:12 Friday 26 March 2021 (16335) Print this report
According to the measurement TF, cross over frequency of temperature and PZT loop is 2mHz (left panels in the attachment)
TF from heater output to error signal of temperature loop (right panels in the attachement) is ~20dB larger than one in klog6824.

We should investigate temperature controller, heater, and thermometer again.
At least thermometer seems to have no response for actuation by heater...
Images attached to this comment
Search Help
×