Reports 1-1 of 1 Clear search Modify search
VIS (PRM)
kouseki.miyo - 20:59 Thursday 26 November 2020 (15540) Print this report
ALIGNED state is available

Now we can move PRM to ALIGNED state as the PRM OpLev was set up today. So we can aligne the PRM suspension for the ghost beam investigation by AOS and IOO team.

  • In the previous work, we did not do the alignement for OpLev table due to no power cable for light source.
  • I cabled the power cable from PRM rack to the OpLev by using the adapter made by AOS.
  • I did not calibrated the OpLev I set up.
  • In order to move ALIGNED state I modified the VIS_PRM.py so that we don't use the boost filter (FM8) for the OLSERVO.
  • Modifications including above one were accepted by SDF as shown in attached figure.
  • In the near future, I will calibrate the OpLev if we will not remove it for long time.
Images attached to this report
Comments to this report:
kouseki.miyo - 21:02 Thursday 26 November 2020 (15542) Print this report

According to Akutsu-san, we had better to prepare the PSD for MISALIGNED state during the investigation. I will do it next week.

kouseki.miyo - 8:31 Friday 27 November 2020 (15547) Print this report
Search Help
×