Reports 1-1 of 1 Clear search Modify search
PEM (Center)
tatsuki.washimi - 9:35 Wednesday 14 October 2020 (15236) Print this report
PEM measurements in power-outage environment

[YokozaWashimi]

We started the PEM measurements in power-outage environment with battery operation, near the OMC chamber.

The reference data during night were not taken because someone kindly turned off our setup without asking us!!!

Images attached to this report
Comments to this report:
tatsuki.washimi - 10:09 Wednesday 14 October 2020 (15237) Print this report

The data during 13:18~13:48 JST has signal, but that of 15:18~15:48 JST has no signal.

Images attached to this comment
tomotada.akutsu - 11:57 Wednesday 14 October 2020 (15238) Print this report

Then, this is also a measurement how "kind" your colleagues are. (and the information would be useful to estimate to what extent we need to prepare of fool-proofing.)

I am not the kind  "someone", but let me put something. When I saw the setup in the evening yesterday, I felt confused why this was left turned on; maybe the memo on the box would not contain sufficient information. Anyway, I myself could estimate what PEM would like to do by taking combination of (a) the information left in my brain and (b) the memo on the box, but I do not think everyone could succeed to combine the information.

tatsuki.washimi - 14:10 Wednesday 14 October 2020 (15239) Print this report

Thank you for your comment.

the memo on the box says "PEM measurement 2020/10/13~14", and we announced this work in KOUTEI-KAIGI and Chief meeting...

tatsuki.washimi - 17:25 Wednesday 14 October 2020 (15241) Print this report

Quick report of the result.

Figure 1 : ASD and coherence for each sensors during today's power outage (13:40~14:10)

Figure 2 : Comparison between today (blue) and yesterday (gray)
(sorry, "VOLT" plot is misreading. the full scale was changed)

 

So quiet!!!

Images attached to this comment
tatsuki.washimi - 17:27 Wednesday 14 October 2020 (15242) Print this report

Measurement is still going.

We will remove them tomorrow.

tomotada.akutsu - 21:48 Wednesday 14 October 2020 (15243) Print this report

Thanks for the interesting plots!

By the way, "VOLT" is measuring what? The Power supply line?

tatsuki.washimi - 8:45 Thursday 15 October 2020 (15245) Print this report

It means the voltage difference between OMC chamber and OMC rack ground

tatsuki.washimi - 13:46 Thursday 15 October 2020 (15247) Print this report

Oct.14 13:40~14:10 vs Oct.15 02:26~0.:56

The difference of "Volt" was real!

Images attached to this comment
shinji.miyoki - 21:48 Thursday 15 October 2020 (15253) Print this report

Same noise enhancement was measured around 100Hz in the vibration noise in the cryostat by Ochi-kun in the past. This origin was identified.