The latest calibration of MCE.
A = 1.10e4 Hz/cnts / f^2 @ISCINF
Procedure:
- lock the IMC
- Measure the TF from ISCINF_EXC to IMC_SERVO_SLOW_DAQ_OUT
- SLOW out can be calibrated by the PZT actuator efficiency 3.0 Hz/V (klog13012)
The latest calibration of MCE.
A = 1.10e4 Hz/cnts / f^2 @ISCINF
- lock the IMC
- Measure the TF from ISCINF_EXC to IMC_SERVO_SLOW_DAQ_OUT
- SLOW out can be calibrated by the PZT actuator efficiency 3.0 Hz/V (klog13012)