* Important Change Information *
As reported earlier, all the segments after 21:13 JST is with DARM2 FM5 (resG2.9)
Input power is different between the segments. The last longest segments was with 2.9W at the IMC OUTPUT.
The first segment was with 2.5W.
We increased to the usual 5W but the lock did not last. So we reduced the power to 4W, and it did not work.
Important Change Information
- As reported earlier, all the segments after 21:13 JST is with DARM2 FM5 (resG2.9)
- Input power is different between the segments, see the main post.
[Supplimental Information]
With 4W or 5W, lock doesn't last more than few minutes. Nakano-kun is suspencting OMC PDs in vacuum are saturating. I checked when a lock was lost with 4W input. See the first attachment. At least for this lockloss, it didn't seem because of the OMC PD saturation. They are saturating very often during TRANSIT_TO_OMCDC (second and third attachments)