Suspension kick for the diagonalization

I made script to kick the suspensions for the diagonalization. I kicked the suspensions at 1252795167.

Script is as follows:

-----------------------------------------------------------------

/users/Commissioning/scripts/kick_suspensions.py
from ezca import Ezca

import time
ezca = Ezca()

for optic in ['BS','SR2','SR3','SRM']:
    ezca['VIS-' + optic + '_IM_COILOUTF_V1_OFFSET'] = 1000
    ezca['VIS-' + optic + '_IM_COILOUTF_H1_OFFSET'] = 5000
    ezca['VIS-' + optic + '_IM_COILOUTF_V1_TRAMP'] = 0.1
    ezca['VIS-' + optic + '_IM_COILOUTF_H1_TRAMP'] = 0.1

    ezca.switch('VIS-' + optic + '_IM_COILOUTF_V1','OFFSET','ON')
    ezca.switch('VIS-' + optic + '_IM_COILOUTF_H1','OFFSET','ON')
    time.sleep(0.1)
    ezca.switch('VIS-' + optic + '_IM_COILOUTF_V1','OFFSET','OFF')
    ezca.switch('VIS-' + optic + '_IM_COILOUTF_H1','OFFSET','OFF')

for optic in ['PR2','PR3','PRM']:
    ezca['VIS-' + optic + '_IM_COILOUTF_V1_OFFSET'] = 1000
    ezca['VIS-' + optic + '_IM_COILOUTF_H1_OFFSET'] = 1000
    ezca['VIS-' + optic + '_IM_COILOUTF_V1_TRAMP'] = 0.1
    ezca['VIS-' + optic + '_IM_COILOUTF_H1_TRAMP'] = 0.1

    ezca.switch('VIS-' + optic + '_IM_COILOUTF_V1','OFFSET','ON')
    ezca.switch('VIS-' + optic + '_IM_COILOUTF_H1','OFFSET','ON')
    time.sleep(0.1)
    ezca.switch('VIS-' + optic + '_IM_COILOUTF_V1','OFFSET','OFF')
    ezca.switch('VIS-' + optic + '_IM_COILOUTF_H1','OFFSET','OFF')           

for optic in ['ITMX','ITMY','ETMX','ETMY']:   
    ezca['VIS-' + optic + '_MN_COILOUTF_V1_OFFSET'] = 8000
    ezca['VIS-' + optic + '_MN_COILOUTF_H1_OFFSET'] = 8000
    ezca['VIS-' + optic + '_MN_COILOUTF_V1_TRAMP'] = 0.1
    ezca['VIS-' + optic + '_MN_COILOUTF_H1_TRAMP'] = 0.1

    ezca.switch('VIS-' + optic + '_MN_COILOUTF_V1','OFFSET','ON')
    ezca.switch('VIS-' + optic + '_MN_COILOUTF_H1','OFFSET','ON')   
    ezca.switch('VIS-' + optic + '_MN_COILOUTF_V1','OFFSET','OFF')
    ezca.switch('VIS-' + optic + '_MN_COILOUTF_H1','OFFSET','OFF')   

 