Reports 1-1 of 1 Clear search Modify search
MIF (General)
yuta.michimura - 21:49 Monday 14 March 2016 (1022) Print this report
Comment to BS re-alignment with picomotors and darkening around the DC PD in the BS chamber.. (981)

On Mar 13, Uchiyama-san changed the PD position so that it faces EXA (see attached photo). The background reduced by ~1/10.
Today, I changed the gain setting from 60 dB to 70 dB. The background now is ~300 counts.

Images attached to this comment
Search Help
×