Reports of 31881
MIF (General)
shinji.miyoki - 22:36 Tuesday 01 July 2025 (34435) Print this report
BNS vs IR Trans Norm

I just compared BNS and IRXY trans Norm for the recent 22 days as Fig.1

  • A very long decreasing trend can be recognized in IRXY_NORM. However, it could be recovered by some adjustments as -19 days, -8 days, and -6 days.
  • The BNS reduction around -7 days seems to correlate with the IRXY trans reduction.

Do you find any more characteristic points?

Images attached to this report
VAC (General)
shinji.miyoki - 22:26 Tuesday 01 July 2025 (34434) Print this report
Comment to X_00 pressure goes up to 6.5e-6 (34431)

N2 increased and decreased as Fig.1

Images attached to this comment
OBS (General)
takahiro.sawada - 19:20 Tuesday 01 July 2025 (34433) Print this report
Comment to Set observing bit (34099)
We turned ON the OBS_INTENT bit at around 17:17 JST on July 1, 2025, following an update to the calibration parameters. There was a lock loss around one minute prior to this due to an earthquake.
VAC (General)
dan.chen - 16:57 Tuesday 01 July 2025 (34432) Print this report
Comment to X_00 pressure goes up to 6.5e-6 (34431)

Plot with "K1:CRY-TEMPERATURE_IX_4K_REF2_4K_HEAD".

Images attached to this comment
VAC (General)
dan.chen - 16:50 Tuesday 01 July 2025 (34431) Print this report
X_00 pressure goes up to 6.5e-6

With Shingo Hido

K1:VAC-PRESSURE_X_00 is going up.

Images attached to this report
Comments to this report:
dan.chen - 16:57 Tuesday 01 July 2025 (34432) Print this report

Plot with "K1:CRY-TEMPERATURE_IX_4K_REF2_4K_HEAD".

Images attached to this comment
shinji.miyoki - 22:26 Tuesday 01 July 2025 (34434) Print this report

N2 increased and decreased as Fig.1

Images attached to this comment
OBS (Summary)
dan.chen - 16:13 Tuesday 01 July 2025 (34430) Print this report
Operation shift summary

Operators name: Takaaki Yokozawa, Hiroshi Takaba, Dan Chen
Shift time: 9-17 (JST)
Check Items:

  • VAC: No issues were found.
  • CRY cooler: No issues were found.
  • Compressor: No issues were found.

IFO was used by CAL group during the day.
After a meeting between Miyoki-san, Ushiba-san, Sawada-san and CAL group, the state is planned to be set to OBSERVING.

CAL (General)
takahiro.yamamoto - 15:41 Tuesday 01 July 2025 (34428) Print this report
Comment to Full calibration measurements (34412)

Finally, I raised CFC_LATCH bit and moved from CALIB_NOT_READY to READY.

Differences in foton filters shown in Fig.1 are related to klog#34424 (K1:CAL-CS_DARM_*) and klog#34425 (K1:CAL-CS_SUM_{MICH,PRCL}_*).
Differences in SDF tables shown in Fig.2 are related to klog#34411 (K1:CAL-MEAS_*) and klog#34426 (K1:CAL-CS_TDEP_*).
There is no change in guardian code and model files as shown in Fig.3 and Fig.4, respectively.

Images attached to this comment
OBS (SDF)
takahiro.yamamoto - 15:28 Tuesday 01 July 2025 (34426) Print this report
Comment to Changes of observation.snap during O4c (34169)

K1CALCS

Channels in JGW-L2314962

It's related to klog#34425.
They were updated based on the latest value of optical gain of DARM and ETMX actuator efficiencies in klog#34424.

Changes were accepted on observation.snap (Fig.1), down.snap (Fig.2), and safe.snap (Fig.3).
Finally, numerical rounding errors were reverted after re-loading observation.snap as shown in Fig.4.

Images attached to this comment
CAL (General)
takahiro.yamamoto - 15:27 Tuesday 01 July 2025 (34425) Print this report
Comment to Full calibration measurements (34412)

MICH and PRCL calibrations were updated by using the values in klog#34423 as follows.
- 1/H_MICH [K1:CAL-CS_SUM_MICH_ERR_FM1]
- A_BS(1) [K1:CAL-CS_SUM_MICH_BS_TM_FM1]
- 1/H_PRCL [K1:CAL-CS_SUM_PRCL_ERR_FM1]
- A_PRM [K1:CAL-CS_SUM_PRCL_PRM_TM_FM1]
Actual changes were shown in Fig.1.

After then line tracking parameters were also updated based on the DARM calibration parameters in klog#34424.
Updated EPICS records can be found in JGW-L2314962.
They were accepted on {observation,down,safe}.snap in klog#34426.

Images attached to this comment
CAL (General)
Shingo Hido - 14:07 Tuesday 01 July 2025 (34424) Print this report
Comment to Full calibration measurements (34412)

I estimated and updated the parameters related DARM (Fig.1).
There is no strange variations (Fig2).

 H_etmxtm = 3.839154366e-14 at 10Hz (0.23% from previous measurements)
 H_etmxim  = 1.535938431e-14 at 10Hz (-1.9% from previous measurements)
 Optical_gain = 2.129473235e12 (0.07% from previous measurements)
 Cavity_pole = 18.127587391 Hz (-0.76% from previous measurements)

Previous parameters are listed in klog#34393.

I investigated the variation in optical gain (specifically, the gain of the sensing function) between 0602 full calibration (klog#34033) and 0630 full calibration (Fig.3).
A decreasing trend was observed from the measurement date (June 2) toward June 20, followed by an increase thereafter.
Since a continuous change can be observed on this time scale (although further data accumulation is needed),
there is a possibility that calibration lines could not only account for statistical fluctuations, but also help suppress systematic errors by correcting for actual variations.

Images attached to this comment
CAL (General)
takahiro.yamamoto - 13:14 Tuesday 01 July 2025 (34423) Print this report
Comment to Full calibration measurements (34412)
Some mistakes were fixed.
1) Wrong measurement file was used for DARM optical gain estimation.
I had used /users/CAL/current/measurements/2025/0630/0920_DARM_OLG.xml in which the measurement was interrupted.
Correct (re-measured) one is /users/CAL/current/measurements/2025/0630/0920_DARM_OLG-1.xml.
I re-analyzed with correct measurement results and H_DARM and 1/H_DARM were updated.
Finally, the difference from previous results is ~3% and a trend of difference will be reported by Hido-kun.

2) Comparison of the actuator efficiency ratio of ETMX and ITMX was done in wrong configuration.
I had compared A_ETMX_TM/A_ITMX_TM(3) in this time and A_ETMX_TM/A_ITMX_TM(0) in the previous measurement.
So I re-evaluated the difference by using A_ETMX_TM/A_ITMX_TM(3) in the previous measurement.

And also comparison between YPcal and XPcal was added in the table.
  Free-swing XPcal Diff. from prev. Figures
H_mich 1.115e+09 +/- 0.0067e+09 N/A -2.8% Fig.1
A_BS(0) 6.180e-11 +/- 0.039e-13 6.000e-11 +/- 0.048e-11 +0.8% (Free-swing) Fig.2-3 (for Free-swing)
A_ITMX(0)/A_BS(0) 0.04956 +/- 0.00038 same as left -0.2% Fig.4-5
A_ITMX(0) 3.063e-12 +/- 0.030e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ITMX(3)/A_ITMX(0) 1.1292 +/- 0.0018 same as left +0.1% Fig.6-7
A_ITMX(3) 3.459e-12 +/- 0.035e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ETMX_TM/A_ITMX(3) 1.1456 +/- 0.0011 same as left -7.9%
-0.9%
Fig.8-9
A_ETMX_TM 3.962e-12 +/- 0.040e-12 3.8574e-12 +/- 0.0033e-12 -0.6% (XPcal) Fig.10-11 (for XPcal)
A_ETMX_IM/A_ETMX_TM 42.1 +/- 2.4 N/A -7.8% Fig.12-13
A_ETMX_IM 1.670e-10 +/- 0.097e-10 1.68e-10 +/- 0.10 -8.7% (XPcal) Fig.14-15 (for XPcal)
H_DARM 2.211e+12 +/- 0.023e+12
2.1246e+12 +/- 0.023e+12
2.2775e+12 +/- 0.0071e+12
2.1875e+12 +/-0.0073e+12
+0.9% (XPcal)
-3.1% (XPcal)
Fig.16-19
=> Fig.1-4 in this post
1/H_DARM 4.522e-13 +/- 0.048e-13
4.709e-13 +/- 0.050e-13
4.391e-13 +/- 0.014e-13
4.572e-13 +/- 0.015e-13
N/A
A_BS(1)/A_BS(0) 1.0423 +/- 0.0048 same as left +0.5% (Free-swing) Fig.20-21
A_BS(1) 6.441e-11 +/- 0.050e-11 6.254e-11 +/- 0.057e-11 +1.3% (Free-swing)
H_MICH 4.628e+08 +/- 0.036e+08 4.767e+08 +/- 0.044e+08 +9.8% (Free-swing) Fig.22-25
1/H_MICH 2.161e-09 +/- 0.017e-09 2.098e-09 +/- 0.019e-09 N/A
A_PRM(1)/A_BS(1) 27.21 +/- 0.12 same as left +1.0% Fig.26-27
M2P FF gain 0.03675 +/- 0.00016 same as left N/A
A_PRM 8.764e-10 +/- 0.078e-10 8.509e-10 +/- 0.086e-10 +2.3% (Free-swing)
H_PRCL 2.057e+09 +/- 0.018e+09 2.118e+09 +/- 0.021e+09 +4.3% (Free-swing) Fig.28-31
1/H_PRCL 4.862e-10 +/- 0.043e-10 4.720e-10 +/- 0.048e-10 N/A
A_ITMX(3)/A_BS(1) 0.05369 +/- 0.00049 same as left N/A
M2D FF gain 18.6 +/- 3.2 same as left N/A
YPCAL/XPCAL N/A 0.98474 +/- 0.00080 -0.2% Fig.5-6 in this post

Images attached to this comment
MIF (General)
takaaki.yokozawa - 12:07 Tuesday 01 July 2025 (34419) Print this report
Comment to Bad alignment? (34418)
The beam position of Yarm back by Yuzurihara-san.

Also, beam spot of OMC trans was not moved, my mis-understanding.
https://gwdet.icrr.u-tokyo.ac.jp/~controls/capture/beamspot/today/latest/
Images attached to this comment
DetChar (General)
shoichi.oshino - 10:50 Tuesday 01 July 2025 (34422) Print this report
Hveto 20250619
The glitch from 60 to 90 Hz has many coincidences with LSC-REFL_PDA3_RF45_I_ERR.
The glitch around 30Hz and 60Hz had coincidence with ASC-DSOFT_P.
The glitch around 70Hz had coincidence with LSC-REFL_PDA2_RF135_I_ERR.
The glitch around 70Hz had coincidence with LSC-AS_PDA1_RF17_Q_ERR.
The glitch from 300 to 400 Hz has many coincidence with ASC-CHARD_P, OMC-QPDV2_YAW, OMC-QPDV1_PIT, ASC-DHARD_P, ASC-REFL_QPDA4_RF11_I_PIT.
DetChar (General)
shoichi.oshino - 10:44 Tuesday 01 July 2025 (34421) Print this report
Hveto 20250617
The glitch around 30Hz and 60Hz had coincidence mainly with ASC-DSOFT_P.
The glitch from 70 to 80 Hz has many coincidence with LSC-POP_PDA1_RF45_Q_ERR.
The glitch around 40Hz had coincidence mainly with LSC-AS_PDA1_RF17_Q_ERR.
The glitch from 300 to 400 Hz has many coincidence with ASC-DHARD_P, ASC_CHARD_P, ASC-REFL_QPDA1_RF45_I_PIT, ASC-AS_QPDA2_DC_SUM.
MIF (General)
takaaki.yokozawa - 9:11 Tuesday 01 July 2025 (34418) Print this report
Bad alignment?
I noticed that the beam position of Yend seemed slightly lower and right, and position of OMC TRANS was slightly high(I don't know).(Fig.1. and Fig.2. and Fig.3.)
I checked the TMSY IR QPD, and it would be shifted in two hours ago and now trying to back to better position.(Fig.4.)
Shift would be due to the earthquake
And detection range was slightly bad compared from stable lock.

Should we add several flag for this event?
Images attached to this report
Comments to this report:
takaaki.yokozawa - 12:07 Tuesday 01 July 2025 (34419) Print this report
The beam position of Yarm back by Yuzurihara-san.

Also, beam spot of OMC trans was not moved, my mis-understanding.
https://gwdet.icrr.u-tokyo.ac.jp/~controls/capture/beamspot/today/latest/
Images attached to this comment
CRY (Alignment survey)
shinji.miyoki - 8:24 Tuesday 01 July 2025 (34417) Print this report
IM Heater adjustment

In last night, I reduced the voltage to heaters for IY and EX IM by ~ 0.3V to reduce the temperature.

Fig.1 shows the relation between IM temperatures and BNS sensitivity. IY temp kept increasing by ~ 1K. On the other hand, the EX temp decreased by 1K.

Fig.2 shows a longer trend. It is difficult to find remarkable relations.

The best BNS sensitivity is 7.35Mpc just before the lock loss. We need calibration?

Images attached to this report
CAL (General)
takahiro.yamamoto - 1:50 Tuesday 01 July 2025 (34416) Print this report
Comment to Full calibration measurements (34412)
Following table is a list of estimated calibration parameters.
  • Each value is estimated based on JGW-L2314903.
  • Estimated values by Pcal and Free-swing are consistent with each other at approximately 3%.
  • Previous full measurement which is the reference of comparison is klog#34033.
  • There is no significant change in all parameters.
  • Numbers in parentheses indicate a number of engaged de-whitening filters.
    • Low-frequency (<1Hz) zero/pole in disengaged de-whitening filters (see also klog#33874) are compensated based on circuit measurement only for ETMX in klog#34229 and klog#34244.
    • So this effect must be considered for other suspensions.
 
  Free-swing XPcal Diff. from prev. Figures
H_mich 1.115e+09 +/- 0.0067e+09 N/A -2.8% Fig.1
A_BS(0) 6.180e-11 +/- 0.039e-13 6.000e-11 +/- 0.048e-11 +0.8% (Free-swing) Fig.2-3 (for Free-swing)
A_ITMX(0)/A_BS(0) 0.04956 +/- 0.00038 same as left -0.2% Fig.4-5
A_ITMX(0) 3.063e-12 +/- 0.030e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ITMX(3)/A_ITMX(0) 1.1292 +/- 0.0018 same as left +0.1% Fig.6-7
A_ITMX(3) 3.459e-12 +/- 0.035e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ETMX_TM/A_ITMX(3) 1.1456 +/- 0.0011 same as left -7.9% Fig.8-9
A_ETMX_TM 3.962e-12 +/- 0.040e-12 3.8574e-12 +/- 0.0033e-12 -0.6% (XPcal) Fig.10-11 (for XPcal)
A_ETMX_IM/A_ETMX_TM 42.1 +/- 2.4 N/A -7.8% Fig.12-13
A_ETMX_IM 1.670e-10 +/- 0.097e-10 1.68e-10 +/- 0.10 -8.7% (XPcal) Fig.14-15 (for XPcal)
H_DARM 2.211e+12 +/- 0.023e+12 2.2775e+12 +/- 0.0071e+12 +0.9% (XPcal) Fig.16-19
1/H_DARM 4.522e-13 +/- 0.048e-13 4.391e-13 +/- 0.014e-13 N/A
A_BS(1)/A_BS(0) 1.0423 +/- 0.0048 same as left +0.5% (Free-swing) Fig.20-21
A_BS(1) 6.441e-11 +/- 0.050e-11 6.254e-11 +/- 0.057e-11 +1.3% (Free-swing)
H_MICH 4.628e+08 +/- 0.036e+08 4.767e+08 +/- 0.044e+08 +9.8% (Free-swing) Fig.22-25
1/H_MICH 2.161e-09 +/- 0.017e-09 2.098e-09 +/- 0.019e-09 N/A
A_PRM(1)/A_BS(1) 27.21 +/- 0.12 same as left +1.0% Fig.26-27
M2P FF gain 0.03675 +/- 0.00016 same as left N/A
A_PRM 8.764e-10 +/- 0.078e-10 8.509e-10 +/- 0.086e-10 +2.3% (Free-swing)
H_PRCL 2.057e+09 +/- 0.018e+09 2.118e+09 +/- 0.021e+09 +4.3% (Free-swing) Fig.28-31
1/H_PRCL 4.862e-10 +/- 0.043e-10 4.720e-10 +/- 0.048e-10 N/A
A_ITMX(3)/A_BS(1) 0.05369 +/- 0.00049 same as left N/A
M2D FF gain 18.6 +/- 3.2 same as left N/A

Images attached to this comment
PEM (Center)
takaaki.yokozawa - 17:01 Monday 30 June 2025 (34415) Print this report
Unti-correlation of the temperature trend between IYA area and outside mine
As shown in Fig.1., I noticed there are unti correlation of the temperature trend between IYA area outside mine.
(At outside, the temperature of noon is high for the outside, but low in IYA area)
This trend continued for a long time(At least more than 100 days), but I cannot find this issue in previous klog.
Some temperature control system worked?
Images attached to this report
OBS (SDF)
Shingo Hido - 16:53 Monday 30 June 2025 (34411) Print this report
Comment to Changes of observation.snap during O4c (34169)

We accepted SDFs related to the cal measurement (klog#34412) in observation.snap, down.snap, and safe.snap (k1calcs).
K1:CAL-MEAS_{CURRENT, LATEST}

Images attached to this comment
CAL (General)
Shingo Hido - 16:52 Monday 30 June 2025 (34412) Print this report
Full calibration measurements

with Dan, YamaT, SawadaT

We performed full measurements for the IFO calibration.
The analysis results will be posted at a later date.

Notes:

The DARM-related measurements were conducted after adjusting the feedforward gains (see klog#34410).
Following the same method as klog#34040 and klog#34033, we performed measurements to observe the changes in DC caused by differences in the number of whitening filter stages for ITMX-TM and BS-TM.

Measurement files (`/users/CAL/current/measurements/2025/0603/0920_*`):
0920_BS_TM_PRCL.xml   : Whitening = 1 stage
0920_BS_TM_PRCL-1.xml  : Whitening = 0 stage
0920_ITMX_TM_DARM_PRFPMI_WFOFF.xml : Whitening = 0 stage
0920_ITMX_TM_DARM_PRFPMI_WFON.xml  : Whitening = 3 stages

Comments to this report:
takahiro.yamamoto - 1:50 Tuesday 01 July 2025 (34416) Print this report
Following table is a list of estimated calibration parameters.
  • Each value is estimated based on JGW-L2314903.
  • Estimated values by Pcal and Free-swing are consistent with each other at approximately 3%.
  • Previous full measurement which is the reference of comparison is klog#34033.
  • There is no significant change in all parameters.
  • Numbers in parentheses indicate a number of engaged de-whitening filters.
    • Low-frequency (<1Hz) zero/pole in disengaged de-whitening filters (see also klog#33874) are compensated based on circuit measurement only for ETMX in klog#34229 and klog#34244.
    • So this effect must be considered for other suspensions.
 
  Free-swing XPcal Diff. from prev. Figures
H_mich 1.115e+09 +/- 0.0067e+09 N/A -2.8% Fig.1
A_BS(0) 6.180e-11 +/- 0.039e-13 6.000e-11 +/- 0.048e-11 +0.8% (Free-swing) Fig.2-3 (for Free-swing)
A_ITMX(0)/A_BS(0) 0.04956 +/- 0.00038 same as left -0.2% Fig.4-5
A_ITMX(0) 3.063e-12 +/- 0.030e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ITMX(3)/A_ITMX(0) 1.1292 +/- 0.0018 same as left +0.1% Fig.6-7
A_ITMX(3) 3.459e-12 +/- 0.035e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ETMX_TM/A_ITMX(3) 1.1456 +/- 0.0011 same as left -7.9% Fig.8-9
A_ETMX_TM 3.962e-12 +/- 0.040e-12 3.8574e-12 +/- 0.0033e-12 -0.6% (XPcal) Fig.10-11 (for XPcal)
A_ETMX_IM/A_ETMX_TM 42.1 +/- 2.4 N/A -7.8% Fig.12-13
A_ETMX_IM 1.670e-10 +/- 0.097e-10 1.68e-10 +/- 0.10 -8.7% (XPcal) Fig.14-15 (for XPcal)
H_DARM 2.211e+12 +/- 0.023e+12 2.2775e+12 +/- 0.0071e+12 +0.9% (XPcal) Fig.16-19
1/H_DARM 4.522e-13 +/- 0.048e-13 4.391e-13 +/- 0.014e-13 N/A
A_BS(1)/A_BS(0) 1.0423 +/- 0.0048 same as left +0.5% (Free-swing) Fig.20-21
A_BS(1) 6.441e-11 +/- 0.050e-11 6.254e-11 +/- 0.057e-11 +1.3% (Free-swing)
H_MICH 4.628e+08 +/- 0.036e+08 4.767e+08 +/- 0.044e+08 +9.8% (Free-swing) Fig.22-25
1/H_MICH 2.161e-09 +/- 0.017e-09 2.098e-09 +/- 0.019e-09 N/A
A_PRM(1)/A_BS(1) 27.21 +/- 0.12 same as left +1.0% Fig.26-27
M2P FF gain 0.03675 +/- 0.00016 same as left N/A
A_PRM 8.764e-10 +/- 0.078e-10 8.509e-10 +/- 0.086e-10 +2.3% (Free-swing)
H_PRCL 2.057e+09 +/- 0.018e+09 2.118e+09 +/- 0.021e+09 +4.3% (Free-swing) Fig.28-31
1/H_PRCL 4.862e-10 +/- 0.043e-10 4.720e-10 +/- 0.048e-10 N/A
A_ITMX(3)/A_BS(1) 0.05369 +/- 0.00049 same as left N/A
M2D FF gain 18.6 +/- 3.2 same as left N/A

Images attached to this comment
takahiro.yamamoto - 13:14 Tuesday 01 July 2025 (34423) Print this report
Some mistakes were fixed.
1) Wrong measurement file was used for DARM optical gain estimation.
I had used /users/CAL/current/measurements/2025/0630/0920_DARM_OLG.xml in which the measurement was interrupted.
Correct (re-measured) one is /users/CAL/current/measurements/2025/0630/0920_DARM_OLG-1.xml.
I re-analyzed with correct measurement results and H_DARM and 1/H_DARM were updated.
Finally, the difference from previous results is ~3% and a trend of difference will be reported by Hido-kun.

2) Comparison of the actuator efficiency ratio of ETMX and ITMX was done in wrong configuration.
I had compared A_ETMX_TM/A_ITMX_TM(3) in this time and A_ETMX_TM/A_ITMX_TM(0) in the previous measurement.
So I re-evaluated the difference by using A_ETMX_TM/A_ITMX_TM(3) in the previous measurement.

And also comparison between YPcal and XPcal was added in the table.
  Free-swing XPcal Diff. from prev. Figures
H_mich 1.115e+09 +/- 0.0067e+09 N/A -2.8% Fig.1
A_BS(0) 6.180e-11 +/- 0.039e-13 6.000e-11 +/- 0.048e-11 +0.8% (Free-swing) Fig.2-3 (for Free-swing)
A_ITMX(0)/A_BS(0) 0.04956 +/- 0.00038 same as left -0.2% Fig.4-5
A_ITMX(0) 3.063e-12 +/- 0.030e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ITMX(3)/A_ITMX(0) 1.1292 +/- 0.0018 same as left +0.1% Fig.6-7
A_ITMX(3) 3.459e-12 +/- 0.035e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ETMX_TM/A_ITMX(3) 1.1456 +/- 0.0011 same as left -7.9%
-0.9%
Fig.8-9
A_ETMX_TM 3.962e-12 +/- 0.040e-12 3.8574e-12 +/- 0.0033e-12 -0.6% (XPcal) Fig.10-11 (for XPcal)
A_ETMX_IM/A_ETMX_TM 42.1 +/- 2.4 N/A -7.8% Fig.12-13
A_ETMX_IM 1.670e-10 +/- 0.097e-10 1.68e-10 +/- 0.10 -8.7% (XPcal) Fig.14-15 (for XPcal)
H_DARM 2.211e+12 +/- 0.023e+12
2.1246e+12 +/- 0.023e+12
2.2775e+12 +/- 0.0071e+12
2.1875e+12 +/-0.0073e+12
+0.9% (XPcal)
-3.1% (XPcal)
Fig.16-19
=> Fig.1-4 in this post
1/H_DARM 4.522e-13 +/- 0.048e-13
4.709e-13 +/- 0.050e-13
4.391e-13 +/- 0.014e-13
4.572e-13 +/- 0.015e-13
N/A
A_BS(1)/A_BS(0) 1.0423 +/- 0.0048 same as left +0.5% (Free-swing) Fig.20-21
A_BS(1) 6.441e-11 +/- 0.050e-11 6.254e-11 +/- 0.057e-11 +1.3% (Free-swing)
H_MICH 4.628e+08 +/- 0.036e+08 4.767e+08 +/- 0.044e+08 +9.8% (Free-swing) Fig.22-25
1/H_MICH 2.161e-09 +/- 0.017e-09 2.098e-09 +/- 0.019e-09 N/A
A_PRM(1)/A_BS(1) 27.21 +/- 0.12 same as left +1.0% Fig.26-27
M2P FF gain 0.03675 +/- 0.00016 same as left N/A
A_PRM 8.764e-10 +/- 0.078e-10 8.509e-10 +/- 0.086e-10 +2.3% (Free-swing)
H_PRCL 2.057e+09 +/- 0.018e+09 2.118e+09 +/- 0.021e+09 +4.3% (Free-swing) Fig.28-31
1/H_PRCL 4.862e-10 +/- 0.043e-10 4.720e-10 +/- 0.048e-10 N/A
A_ITMX(3)/A_BS(1) 0.05369 +/- 0.00049 same as left N/A
M2D FF gain 18.6 +/- 3.2 same as left N/A
YPCAL/XPCAL N/A 0.98474 +/- 0.00080 -0.2% Fig.5-6 in this post

Images attached to this comment
Shingo Hido - 14:07 Tuesday 01 July 2025 (34424) Print this report

I estimated and updated the parameters related DARM (Fig.1).
There is no strange variations (Fig2).

 H_etmxtm = 3.839154366e-14 at 10Hz (0.23% from previous measurements)
 H_etmxim  = 1.535938431e-14 at 10Hz (-1.9% from previous measurements)
 Optical_gain = 2.129473235e12 (0.07% from previous measurements)
 Cavity_pole = 18.127587391 Hz (-0.76% from previous measurements)

Previous parameters are listed in klog#34393.

I investigated the variation in optical gain (specifically, the gain of the sensing function) between 0602 full calibration (klog#34033) and 0630 full calibration (Fig.3).
A decreasing trend was observed from the measurement date (June 2) toward June 20, followed by an increase thereafter.
Since a continuous change can be observed on this time scale (although further data accumulation is needed),
there is a possibility that calibration lines could not only account for statistical fluctuations, but also help suppress systematic errors by correcting for actual variations.

Images attached to this comment
takahiro.yamamoto - 15:27 Tuesday 01 July 2025 (34425) Print this report

MICH and PRCL calibrations were updated by using the values in klog#34423 as follows.
- 1/H_MICH [K1:CAL-CS_SUM_MICH_ERR_FM1]
- A_BS(1) [K1:CAL-CS_SUM_MICH_BS_TM_FM1]
- 1/H_PRCL [K1:CAL-CS_SUM_PRCL_ERR_FM1]
- A_PRM [K1:CAL-CS_SUM_PRCL_PRM_TM_FM1]
Actual changes were shown in Fig.1.

After then line tracking parameters were also updated based on the DARM calibration parameters in klog#34424.
Updated EPICS records can be found in JGW-L2314962.
They were accepted on {observation,down,safe}.snap in klog#34426.

Images attached to this comment
takahiro.yamamoto - 15:41 Tuesday 01 July 2025 (34428) Print this report

Finally, I raised CFC_LATCH bit and moved from CALIB_NOT_READY to READY.

Differences in foton filters shown in Fig.1 are related to klog#34424 (K1:CAL-CS_DARM_*) and klog#34425 (K1:CAL-CS_SUM_{MICH,PRCL}_*).
Differences in SDF tables shown in Fig.2 are related to klog#34411 (K1:CAL-MEAS_*) and klog#34426 (K1:CAL-CS_TDEP_*).
There is no change in guardian code and model files as shown in Fig.3 and Fig.4, respectively.

Images attached to this comment
MIF (General)
hirotaka.yuzurihara - 16:47 Monday 30 June 2025 (34414) Print this report
Comment to Blasting of Tunnel construction unlocked IFO (33601)

Here is a update of the 8 blasting between 2025/06/18 03:22:49 UTC and 2025/06/26 13:59:52 UTC. The plots are available at DAC wiki.

  • 6 times: lockloss occurred from OBSERVATION state
    • The the amplitude of the seismic motion was large enough to make the lockloss.
    • Even though the bomb size was 6.2 kg (which is relatively small), the lockloss occurred. I guess this is because the blasting point is approaching to the KGRA tunnel.
  • 2 times: interferometer was in down state or during lock acquisition.
OBS (Summary)
dan.chen - 16:23 Monday 30 June 2025 (34413) Print this report
Operation shift summary

Operators name: Takaaki Yokozawa, Hiroshi Takaba, Dan Chen
Shift time: 9-17 (JST)
Check Items:

  • VAC: No issues were found.
  • CRY cooler: Temperature of EX 50_REFBRT went down by 6K between 6/29 16:00 and 6/30 10:00. No other movement.
  • Compressor: No issues were found.

IFO was used by CAL group during the day.
After their works, the state was set to be CALIB NOT READY. 

MIF (General)
takafumi.ushiba - 15:12 Monday 30 June 2025 (34410) Print this report
Tuning of feedforward gains

[Dan, Hido, Sawada, Yamamoto, Ushiba]

Abstract:

Gain tuning of MICH2DARM, MICH2PRCL, and PRCL2DARM FF were performed.
All update seems fine.

Detail:

First, calibration team measured the actuator efficiency ratio of MICH2PRCL/PRM2PRCL.
The new value was 0.03675 and FM10 of LSC_MICHFF2 filter bank was updated.
Figure 1 and 2 show the filters before and after the update, respectively.

Then, I measured MICH2DARM coupling with several FF gain to tune the MICH2DARM FF gain (fig3).
MICH FF performance seems better when MICH FF gain is 0.98, so I implemented the gain of 0.98 at FM10 of LSC_MICHFF1 filter bank.
Figure4 and 5 show the filters before and after the update, respectively

After that, I measured PRCL2DARM coupling with several FF gain (fig6).
Though the low frequency performance is worse than before, the best gain seems 0.8, so I update FM1 of LSC_PRCLFF1 filter bank.
Figure 7 and 8 show the filters before and after the update, respectively.
For further tuning of PRCL2DARM FF, we may need to modify the filter shape and it takes a time, so I didn't performit today.
Since current PRCL2DARM noise budget cased on the today's measurement is low enough (fig9), the current situation should be fine.

Since all updates are done by overwriting current FF filters, no SDF change happens.

Images attached to this report
MIF (General)
takaaki.yokozawa - 13:08 Monday 30 June 2025 (34409) Print this report
Sensitivity study 250630
I tried to search the several correlation phenomena between KAGRA detection range and auxiliary channels (I used phenomena since I checked not only time series but also spectrogram and whitened spectrogram.)
Still I cannot find the critical reason, but I found the 80 Hz peak appeared in about every 7-8 min and high frequency in 120 Hz and 160 Hz
Images attached to this report
DGS (General)
shinji.miyoki - 8:53 Monday 30 June 2025 (34408) Print this report
Comment to Timing synchronization was lost on ITMX and ITMY (34320)

5 days have passed since the last recovery. No problem has happened until now. So, the heat up inside the 19-inch rack could be the reason.

Search Help
×

Warning

×