IOO (Laser Bench)
osamu.miyakawa - 12:45 Friday 29 March 2024 (29008) Print this report
Removing HWP at PSL

I removed the half-wave plate (HWP) that changed the polarization and was located between the shutter and the periscope at the very end of the PLS table (before Fig.1, after Fig.2).

Now the polarization is back to S.

I left the HWP on the PSL table just the side of the shutter. I saw a beam to RFAM and was careful not to block it by the HWP.

By the way, I noticed that the aperture of the HWP was only .5 inch. The beam diameter looked more than 1cm on the sensor card. Was there no clipping?

Images attached to this report
Comments to this report:
tomotada.akutsu - 12:36 Friday 29 March 2024 (29011) Print this report

> I noticed that the aparture of the HWP was only .5 inch. Beam diameter looked more than 1cm on the sensor card. Was there no clipping?

Some of us have already known it (for example 28776). We have been worried about this, and so sometimes this point was discussed in some morning meetings. In any case, please prepare the better HWP for the next case.