Reports 1-1 of 1 Clear search Modify search
MIF (ITF Control)
masayuki.nakano - 21:29 Wednesday 17 June 2020 (14564) Print this report
Comment to PRC length measurement (14555)

PRM calibration procedure

- Lock PRY
- Measure the transfer functions from BS(PRM)_ISCINF to PRCL_IN1 (shown in first attached figure.)
- Take ratio of them (top left of second attached figure)

The ratio of PRM actuator efficiency to BS one is ~47.5, and the BS actuator efficiency is 6.28e-11/f^2 m/ct (klog14557). So the actuator efficiency of PRM is 2.983e-11/f^2.

This number is inconsistent with the number got in Febrary(13020). Did we change anything...?

Images attached to this comment