[Nakano, Akutsu, Michimura]
We have calibrated PRM actuator using PRY cavity.
Measured actuation efficiency ratio between PRM and BS was PRM/BS = 16.8.
Using Ogaki-kun's cablibration of BS actuator (klog #11754), this suggests that PRM actuation efficiency is 1.0e-9/f^2 m/cnt in the free mass region.
Method:
1. Locked PRY (PRM-ITMY cavity) with a UGF of 60 Hz.
2. Measured the transfer function from K1:LSC-PRCL2_OUT to K1:LSC-PRCL1_IN1 when PRY is locked with BS or PRM (see attached).
3. Did the weighted fitting of the transfer functions with Act/freq^2. Act for BS was 3.1e-2 and that for PRM was 5.2e-1 (see attached).
4. According to Ogaki-kun's calibration in Nov 25, A_BS = 6.02e-11/f^2 m/cnt (I think this is with a high power coil driver and we are still using a high power coil driver for BS TM). Therefore,
A_PRM = 5.2e-1/3.1e-2 * 6.02e-11/f^2 m/cnt = 1.0e-9/f^2 m/cnt.
Discussion:
PRM TM also uses a high power coil driver. Theoretical actuation efficiencies for PRM and BS OSEM are 0.129 N/A and 0.0138 N/A, respectively, and the mirror masses of PRM and BS are 10.8 kg and 18.9 kg, respectively [1]. This suggests that actuation efficiency ratio will be
PRM/BS = (0.129 N/A / 10.8 kg) / (0.0138 N/A / 18.9 kg) = 16.4 .
So, the ratio is consistent with the measurement.
[1] https://iopscience.iop.org/article/10.1088/1361-6382/aa90e3
Next:
- Calibrate PRCL and check the displacement sensitivity and the residual RMS during locking