Reports 1-1 of 1 Clear search Modify search
PEM (Center)
takaaki.yokozawa - 0:31 Thursday 19 March 2020 (13677) Print this report
Acoustic injection to REFL table

I applied the acoustic injection to REFL table

Injection : 1268580000 - 1268580110

DAC output 10000count 60-1000Hz uniform

silent : 1268579700 - 1268579900

Images attached to this report
Comments to this report:
tatsuki.washimi - 9:34 Thursday 19 March 2020 (13683) Print this report

I analized this data.

Background : 1268579800-1268579900
Injected : 1268579970-1268580070
Main channel : K1:CAL-CS_PROC_C00_STRAIN_DBL_DQ
Monitor channel : K1:PEM-SENSOR_BOOTH_IMC_BNC3_OUT_DQ

The 280 Hz bump and the 520Hz bump are well reduced. The 360Hz bump is still.

Images attached to this comment
tatsuki.washimi - 11:23 Thursday 19 March 2020 (13684) Print this report

From a discussion with ASO-san and Nakano-san, the slope in 50-300 Hz may be due to that FF was not applied during this time.