at Mitaka
I measured the actuator efficiency of coil for TMS-VIS with the same setup as yesterday's (8138).
I used the voltage which was displayed on the signal generator's screen in the result of yesterday's measurement, though they adopted estimated output in the previous measurenent (5014).
So I modified the voltage to adjust the conditions. (estimated voltage is adopted as the same manner as klog5014)
Attached plots show the results of the coil actuator labeld No.13, 15, and 16.
No.14 was broken due to wire's disconnection...
[Results] (all value have same unit [N/V])
No.13 : 0.0263
No.15 : 0.0264
No.16 : 0.0266
These results seem compatible with the previous results (5014).
[Note]
Glue remained on the coil wire...