Reports 1-1 of 1 Clear search Modify search
VIS (BS)
enzo.tapia - 9:31 Friday 18 January 2019 (7694) Print this report
Comment to BS Service During Vent (7675)

Here are the plots and data for the F0 LVDT and the F0 FR (Fishing Rod).

Position of keystone [mm] F0 LVDT raw output [cnt]
11.6 4175
12.5 3220
13.5 1780
14.5  930
15.5 77
16.5 -870
17.5 -1825
18.5 -2720
19.5 -3160
20.5 -3690

Calibration factor: 1/(896.23[cnt/mm]) = 1.116[um/cnt]

Offset: 870[cnt]

I have set the calibration factors and the offset into the digital system and I took a Snapshoot

Steps in command line [-] F0 LVDT [µm]
-1400000 -2552.9
-1200000 -2242.7
-1000000 -1774.1
-750000 -1041.0
-500000 -214.2
-300000 464.2
-100000 1096.8
0 1237.4
100000 1494.0
300000 1991.7
500000 2442.5
750000 2549.6
1000000 4264.5
1200000 4642.8
1400000 4946.3


It can be seen that the F0 can be moved using the FR by 2.5mm downwards and 4.9 mm upwards.