kohei.yamamoto - 19:12 Monday 12 November 2018 (6965) Print this report
dB for the nominal modulation index

Fujii, Kohei Yamamoto

We measured needed Voltage[dBm] on EOM to generate the nominal modulation indices.
The data is shown on the table I attached. "with RF amp" means the combination of 17[dB] RF amp and 7[dB] attenuator.
Colored values indicate the necessity of RF amp. 
Figures are attached in the same order as that in the table.

Images attached to this report
Comments to this report:
kohei.yamamoto - 12:46 Tuesday 20 November 2018 (7048) Print this report

With the fix I presented in 7047, again I did the same measurement.
The figure is the new version with a minor correction.
This is slightly different from the original data. 

As I report in 6971, the gain of combination of RF amp & attenuator is higher than the spec about 3dB.
 

Images attached to this comment