yoshinori.fujii - 21:04 Friday 11 May 2018 (5002) Print this report
BF-LVDT calibration

  The BF-LVDTs for ITMX are calibrated and the results are in the attached. Briefly, the calibration factors are summarized as the following:

BF-LVDT calibration factor [um/count]
V1 -0.354
V2 -0.367
V3 0.385
H1 0.371
H2 0.382
H3 0.387


  The attached also includes the measured noise spectra of these BF-LVDTs. The noise floor seems to become larger if the primary coil is off the center of the linear range. At the worst case, the noise floor becomes larger by on order of magnitude (we will not use the BF-LVDT at such position though).

[REMEMBER ME]
-DOUBLE-CHECK THE POLALIZATION AFTER INSTALLATION
  --I have tagged the parts which supports BF-LVDT coils about the information of the calibration axes, I hope it will help the installation.