Reports 1-1 of 1 Clear search Modify search
AOS (Baffles & Beam dumps)
kouseki.miyo - 8:33 Thursday 22 February 2018 (4246) Print this report
Calibration of photo sensor for WAB cooling test

Simon, Miyo

Today(yesterday), we calibrate all photo sensor changing the power source condition; Agilent F3630A or OSEM Satelite box.

Tomorrow(today), We will install photo sensor on the WAB.

Images attached to this report
Comments to this report:
kouseki.miyo - 9:56 Thursday 22 February 2018 (4249) Print this report

On klog:4231, S2 signal is different from others. We guess that this difference was caused by the reflector surface difference . The pair of sensor head and reflector surface is shown in attached file.
So, we change the pair of sensor. Green is correct S2 pair (S2 and +X surface). Red is correct S3 pair (S3 and -Z surface ). Purple is S3 and +X surface which is we thought bad surface. 
Even we change surface pair, S3 signal was not changed. So We guess S2 sensor head has some problem. 