We re-calibrated MC oplevs again.
Since mirror angle d_theta is:
Regular configuration d_theta = dx / (2*lever arm length)
Folded configuration d_theta = dx / (8*lever arm length),
calibration factor [urad/cts] = slope [mm/cts] / (2*lever arm length [m]) * 1000
MCE - lever arm length 0.95m
PIT 145 [urad/cts]
(slope 0.28 [mm/cts], linear range ~ 0.4 mm)
YAW 145 [urad/cts]
(slope 0.28 [mm/cts], linear range ~ 0.5 mm )
MCI - lever arm 0.86m
PIT 312 [urad/cts]
(slope 0.54 [mm/cts], linear range ~ 0.85 mm )
YAW 286 [urad/cts]
(slope 0.49 [mm/cts], linear range ~ ? micrometer range out)
MCO (folded) - lever arm 0.85 m
PIT 102 [urad/cts]
(slope 0.69 [mm/cts], linear range ~ 1 mm )
YAW 97 [urad/cts]
(slope 0.66 [mm/cts], linear range ~ 0.75 mm)
Lever arm lengths were taken from klog #570.
I also upload calibrated spectra (it was taken at a noisy work timing, with people walking around MCF).