[Takahashi, Sato, Okutomi]
BF GAS LVDT calibration
- Sensor efficiency: (346 +/- 14) V/m (see attached figure)
- LVDT outputs are single-ended obtained with the Dsub=>BNC Converter (as reported in #2489)
Release Sensor check
- We confirmed that the Release Sensors work correctly
- The LEDs on the indicator box are lighted when a mass is sitting
- Cabling work of the Release Sensors on the stages of DP and BF is also completed
Adjustable range check with Fishing Rod
-
Height of the keystone is adjustable only in the range of ~ 2 mm
- We expected the range to be ~ 10 mm, so this is too small
- We could NOT let the Dummy Payload (DP) sit on the Release Sensors with the Fishing Rod height adjustment
- This could happen because blade springs on the Fishing Rod are designed for the optimal weight of the BF as ~ 50 kg, about a quarter of that of the DP
- We'll add more blade springs parallel to the existing ones which are now in Mitaka
Photos >>> https://goo.gl/photos/aiRxGj4ty26uxgkx8