Reports 1-1 of 1 Clear search Modify search
VIS (IX)
koji.nakagaki - 19:14 Tuesday 06 July 2021 (17405) Print this report
IX Pico motor driver was broken and replaced.
[Miyo,Nakagaki]  2021/7/6 9:30-10:30

The ITMX pico motor driver was not responding on the MEDM, so we checked it on site and found that the device had failed.
(The power was on, but the power indicator and LAN port lights were off.)

It was replaced with a spare part and recovered.

Faulty device will be sent to the manufacturer for investigation & repair.
Images attached to this report
Comments to this report:
satoru.ikeda - 12:50 Wednesday 07 July 2021 (17420) Print this report

[Ikeda, Miyo-san, Nakagaki-san]
I checked the Picomotor log and there is evidence that it was working properly on 06/25/2021 at about 8:45am.
Based on the log, we think it was corrupted between 6/25 and 7/5.

Next, I smelled a smell from the broken Picomotor driver and decided to disassemble it to check it out instead of sending it in for repair.
JGW-S180872 S/N: 8742-39212
[Status.]
The left side pin of the two DC jacks that are not in use is bent.
The left side of the DC connector shows signs of having been burned near its base.
Based on the above, it is highly likely that the DC connector was deformed, or that a capacitor or resistor at the root of the DC connector was burned.

 

Images attached to this comment