Reports 1-1 of 1 Clear search Modify search
VIS (IX)
satoru.ikeda - 10:43 Thursday 15 April 2021 (16480) Print this report
Comment to Height adjustment of ITMX (16421)

All models of k1ix1 suddenly died while driving F1 GAS-Stepper motor of ITMX.

I learned from yamamoto-san and used the following command to restore all models including the iop server.
/opt/rtcds/userapps/release/cds/common/scripts/restart_iop_model.sh

The current value is currently running at 26A/30A, and there is a possibility that the upper limit has been exceeded by driving Stepper.
If the problem occurs frequently, it is necessary to strengthen the power supply.