Setup of BF

[Takahashi, Sato, Horiuchi]

  1. Set the height of BF to 70mm.
  2. Released the DP.
  3. Released the kyestone of BF.
  4. Replaced the signal generator (SG) from the commertial one to the embedded board. Original SG was set to 7.7Vp-p (5.8Vp-p for each driver) in amplitude. New SG was set to 7.0Vp-p. 
  5. Calibrated the LVDT in BF (see plot #1).
  6. Tuned the BF at the mid point (41mm in height) of FR (see plot #2). Base weight was 202.7kg.
  7. Checked the range of FR (see plot #3).
  8. Fixed the keystone.
  9. Set the height of BF to 60mm.
Images attached to this report
Comments to this report:
kouseki.miyo - 11:59 Wednesday 20 January 2021 (15788) Print this report

I updated K1:VIS-ETMY_BF_LVDTINF_GAS_GAIN from -0.883 [um/count] to -0.848 [um/count] according to your measurement, after that I accepted this modification in SDF as you can see in the attached file.

Images attached to this comment
kouseki.miyo - 12:09 Wednesday 20 January 2021 (15789) Print this report

I updated K1:VIS-ETMY_BF_LVDTINF_GAS_GAIN from -0.883 [um/count] to -0.848 [um/count] according to your measurement, after that I accepted this modification in SDF as you can see in the attached file.
The old one was calibrated by Okutomi-san in klog#2499.