Hirata, Yano, Akutsu
Finished the WAB moving to the position in saving mode (Fig 1 and 2).
- Fixed the WAB suspended part with the triangle protectors. A thermometer on the WAB suspension structure was fixed to a screw hole to be used for the protector, so the thermometer was detached.
- Detach the 6N Al heat links and the adapter block to the cooling bar.
Then, we started the WAB moving.
- We adjusted the height of two foot blocks to be set for the WAB when it is in saving mode before hunging the WAB.
- We slightly hung the WAB up and detach the clamps (slightly one by one) that are fixing the WAB base plate.
- We moved the WAB to the position in saving mode.
FInally I pulled pins of thermometers for the WAB one by one out from the burndy connector, and pushed them into a PEEK connector (Fig 3 and 4). Then I and Ushiba-kun checked their connectivirty; that was ok.
Note that the adaptor block for the 6N Al heat links is the old one (Fig 5 and 6). We need to replace it. I looked for the new one but could not find it in the office... so it maybe in the tunnel?