We performed the acoustic injection to the PSL room.
GPSTime : 1269111350 - 1269111500
60-1000Hz uniform 10000DAC count
We performed the acoustic injection to the PSL room.
GPSTime : 1269111350 - 1269111500
60-1000Hz uniform 10000DAC count
This result is not REFL table, but PSL room. So we didn't confirm the effect of reducing scattered light at REFL table.
Background : 1269111510 - 1269111610
Injected : 1269111395 - 1269111495
Main channel : K1:CAL-CS_PROC_C00_STRAIN_DBL_DQ
PEM channel : K1:PEM-MIC_PSL_TABLE_PSL4_Z_OUT_DQ
The acoustic noise level over 300 Hz in PSL room is lower than the current sensitivity about with factor 3.
I perform the NInjA analysis to accelerometer in the PSL room.
1. 225.3Hz : on the PSL table (ACCs shape is suitable)
2. 262.4Hz : ACC on the periscope at exit point
3. 292Hz : on the PSL talbe (ACC3 is largest)
4.315Hz : ACC on the periscope at eit point
5. 335Hz : PSL mic
6. 370Hz : on the PSL table
7. 380 Hz : Periscope for the green fiber (coherence with ALS CARM)