Result report: X arm bounce CR up to 8th, f1 up to 2nd, f2 up to 2nd minus and zero data was avoided by data-min(OMC)*1.005 peak amp [mW], freq [MHz], width [MHz] == Carrier == CA0_1: 9.2273 -0.29369 0.21366 CA0_2: 9.6577 208.8569 0.20736 CA1: 0.37742 64.1126 0.1867 CA2: 0.62695 128.9081 0.19462 CA3: 0.25628 193.873 0.21074 CA4: 0.10377 49.1326 0.22692 CA5: 0.0028023 113.9115 0.038156 CA6: 0.024344 178.7663 0.21478 CA7: 0.011652 34.0072 0.25845 CA8: 0.0060424 98.5293 0.27802 == 45 MHz SB == USB0: 0.049988 44.6436 0.16245 USB1: 0.013997 109.1804 0.19781 USB2: 0.0034115 174.2035 0.16895 LSB0: 0.05367 163.725 0.1977 LSB1: 0.017438 19.1887 0.21731 LSB2: 0.0042733 83.4195 0.46674 == 17 MHz SB == usb0: 0.040659 16.5705 0.15564 usb1: 0.013469 80.998 0.18658 usb2: 0.0026199 145.938 0.27311 lsb0: 0.041275 191.9633 0.15783 lsb1: 0.012933 47.2651 0.1845 lsb2: 0.014474 113.7147 0.30417 Mode matching ratio (P_CR0th vs P_CRall) is 87.0136%. Ratio between P_CR0th vs P_CR2th is 3.997%. Fitted FSR is 209.1505 MHz. Finesse is 561.9829 with standard deviation of 16.4576 Elapsed time is 189.343322 seconds.