Reports 1-1 of 1 Clear search Modify search
CAL (General)
wataru.ogaki - 8:19 Tuesday 19 November 2019 (11633) Print this report
measurement of ratios between actuator efficiencies
This morning I tried measurement of actuator efficiency ratios with FPMI.

Successfully measured (pic.1):
A_{ETMY}/A_{ITMY} = (DARM1_IN1 / ETMY_TM_LOCK_L_EXC) / (DARM1_IN1 / ITMY_TM_LOCK_L_EXC)
=6.021 (this is consistent with the past value measured by Yamamoto-san in 11252)

However, when measuring for
A_{ITMY}/A_{BS} = (MICH1_IN1 / ITMY_TM_LOCK_L_EXC) / (MICH1_IN1 / BS_TM_LOCK_L_EXC),
I found that only A_{BS}'s slope seems strange (pic.2).
I suspected existence of some filters in the path of the excitation signals, but there seems no filters.

So today I couldn't complete propagation of actuator efficiency from BS to ETMY.
I want to measure them again on another day, including calibration of absolute value of A_{BS}.
Images attached to this report
Search Help
×

Warning

×