VAC (IX)
nobuhiro.kimura - 9:03 Thursday 27 January 2022 (19590) Print this report
Result of Q-mass measurement at IX and EX
When: 27/Jan./2022
Who:Kimura and Yasui
Task:
We are started Q-mass measurement at IX and EX.
Please see attached photos.
Phots show only raw date including noise of caused by ionization in Q-mass.
It needs to analyze raw data to reject the noise.
Images attached to this report
Comments to this report:
nobuhiro.kimura - 8:20 Friday 28 January 2022 (19610) Print this report

When: 27/Jan./2022
Who: Kimura and Yasui
Task:
We measured partial pressure of IXC and EXC by manual mode of Q-mass analzer.

Results are as follows;

  IX EX
Tortal Pressure (Pa) 3.1 x 10^-4 3.1 x 10^-4
He (Pa) 2.0 x 10^-10 7.7 x 10^-10
H2O (Pa) 7.5 x 10^-6 2.7 x 10^-4
O2 (Pa) 4.8 x 10^-8 1.7 x 10^-6
N2 (Pa) 2.1 x 10^-7 7.6 x 10^-6
Ratio of O2/N2 (%) 23 22

We confirmed main component of redisual gas in the cryostat was H2O.

We continue to measure and analyze redisual gas in the cryostat.

nobuhiro.kimura - 12:07 Friday 04 February 2022 (19676) Print this report

When: 3/Feb./2022
Who: Kimura and Yasui
Task:
We measured partial pressure of IXC and EXC with manual mode of Q-mass analyzer, again.

Results are as follows;

  IX EX
Tortal Pressure (Pa) 3.1 x 10^-4 2.9 x 10^-4
He (Pa) 1.6 x 10^-10 8.8 x 10^-10
H2O (Pa) 6.5 x 10^-6 2.6 x 10^-4
O2 (Pa) 4.1 x 10^-8 1.7 x 10^-6
N2 (Pa) 2.2 x 10^-7 8.6 x 10^-6
Ratio of O2/N2 (%) 20 20

We confirmed main component of redisual gas in the cryostat was H2O.

We continue to measure and analyze redisual gas in the cryostat.