When: 27/Jan./2022
Who: Kimura and Yasui
Task:
We measured partial pressure of IXC and EXC by manual mode of Q-mass analzer.
Results are as follows;
IX | EX | |
Tortal Pressure (Pa) | 3.1 x 10^-4 | 3.1 x 10^-4 |
He (Pa) | 2.0 x 10^-10 | 7.7 x 10^-10 |
H2O (Pa) | 7.5 x 10^-6 | 2.7 x 10^-4 |
O2 (Pa) | 4.8 x 10^-8 | 1.7 x 10^-6 |
N2 (Pa) | 2.1 x 10^-7 | 7.6 x 10^-6 |
Ratio of O2/N2 (%) | 23 | 22 |
We confirmed main component of redisual gas in the cryostat was H2O.
We continue to measure and analyze redisual gas in the cryostat.
When: 3/Feb./2022
Who: Kimura and Yasui
Task:
We measured partial pressure of IXC and EXC with manual mode of Q-mass analyzer, again.
Results are as follows;
IX | EX | |
Tortal Pressure (Pa) | 3.1 x 10^-4 | 2.9 x 10^-4 |
He (Pa) | 1.6 x 10^-10 | 8.8 x 10^-10 |
H2O (Pa) | 6.5 x 10^-6 | 2.6 x 10^-4 |
O2 (Pa) | 4.1 x 10^-8 | 1.7 x 10^-6 |
N2 (Pa) | 2.2 x 10^-7 | 8.6 x 10^-6 |
Ratio of O2/N2 (%) | 20 | 20 |
We confirmed main component of redisual gas in the cryostat was H2O.
We continue to measure and analyze redisual gas in the cryostat.