Reports 1-4 of 4 Clear search Modify search
DGS (General)
yukitsugu.sasaki - 15:47 Thursday 24 March 2016 (1187) Print this report
conlog setup
We have finished the setup of KAGRA conlog.
Let's use the conlog at KAGRA.
http://k1script/conlog/conlog_web.pl
Images attached to this report
DGS (General)
yukitsugu.sasaki - 0:36 Friday 26 February 2016 (837) Print this report
Install of Thermo-hygrometer in center area.
[Sasaki,Kokeyama]

We installed thermo-hygrometers and one base station in X end,Y end and center area.

EPICS displays the temperature and humidity of X end, Yend and center area.

We installed probe on PSL room, X wall, Y wall, and warehouse.

We can confirm example of these data.(See the attached figure.)

Thermo-hygrometers script is currently running with SuperVisor on k1script.

Detail of SuperVisor is described in DGS blog.
Images attached to this report
Comments to this report:
keiko.kokeyama - 18:44 Friday 26 February 2016 (853) Print this report

EXA, EYA and EX1 are also installed today. They will be added on the EPICS channels.

DGS (General)
yukitsugu.sasaki - 11:38 Friday 29 January 2016 (702) Print this report
Comment to IMC locked automatically (698)
This is brief description of figure.

Guardian for IMC LOCK consist of many state.

First, INIT STATE initial state at startup. Currently, this state does nothing, just go to Down STATE.

DOWN STATE reset all configuration.

ZERO_OFFSETS STATE set zero any offset before engaging servo.

FIND_RESONANCE STATE move mirror to find resonance condition before engaging slow servo.

INCRESE_GAIN is currently empty, but will be used to increase gain.

finally, IMC_LOCKED STATE check for lockloss.
Images attached to this comment
DGS (General)
yukitsugu.sasaki - 20:56 Thursday 28 January 2016 (698) Print this report
IMC locked automatically
We have completed automatically IMC LOCK with Guardian.

Guadian operate mirror and Lazer PZT.

DGS Blog is described in detail.
Images attached to this report
Comments to this report:
yukitsugu.sasaki - 11:38 Friday 29 January 2016 (702) Print this report
This is brief description of figure.

Guardian for IMC LOCK consist of many state.

First, INIT STATE initial state at startup. Currently, this state does nothing, just go to Down STATE.

DOWN STATE reset all configuration.

ZERO_OFFSETS STATE set zero any offset before engaging servo.

FIND_RESONANCE STATE move mirror to find resonance condition before engaging slow servo.

INCRESE_GAIN is currently empty, but will be used to increase gain.

finally, IMC_LOCKED STATE check for lockloss.
Images attached to this comment
Search Help
×