With Hido Shingo
We performed the fine adjustment for the Pcal-X alignment with the ALIGNED state of the ETMX suspension.
We used PCAL_EX1(in Tx module) and PCAL_EX2(in EXA chamber) picomotors to adjust the path 1 so that the paths passed through the design position on the ETM and the center of the RxPD.
We left the Pcal-X laser ON, and the Pcal-X guardian state at OFS_CLOSED_LOW_POWER.
Attached files are showing beam positions on ETM and RxPD.
The total picomotor steps we requested are:
The final optical efficiencies were (quick measurement):
These are similar to klog28820.
By analyzing the recent TCam images, I identified the mirror center of ETMX, ITMX, and ITMY. When the ETMY or GreenY is ready, we will try to take the ETMY image and identify the mirror center.
Details
[Kimura and M. Takahashi]
To identify the source of the liquid material that entered the EYA tank, three filters of the air compressor and the inside of the connection piping of the dry pumps for vacuum pumping were visually inspected.
As a result of the visual inspection, no traces of liquid material were found inside the filters of the air compressor and the connecting piping of the dry pumps for vacuum pumping.
The smell inside the filter and piping was confirmed by nose, but the same irritating smell observed in the EYA tank was not found.
Three filters of the air compressor were replaced with new ones because more than two years had passed since the last replacement.
[Kimura and Ueda (SKS) ]
On the afternoon of 18 Apr., Ueda-san and me performed a vacuum leak test for new pressure gage of GVetmx.
The results of the vacuum leak test confirmed that the new pressure gage of GVetmx did not leak more than 1x10^-12 Pam^3/s.
[Kimura, and M. Takahasshi]
On the morning of 25 Apr., M. Takahashi-san and me performed a vacuum leak test for new pressure gage of GVetmy.
(See attached photos 1~2)
The results of the vacuum leak test confirmed that the new pressure gage of GVetmy did not leak more than 1x10^-12 Pam^3/s.
After the leak test, we set up protection panel in front of the gage.
(See attached photo 3)
The protection panel for the gage of GVetmx was set up, too.
[YokozaWashimi, Ishikawa, Ozaki, Sudo]
We performed hammering tests (vertical tapping) for the OMC in-vac table.
The tapped points are the table, the base plate, and the ground near the stack1 or 2.
Yamamoto, Tanaka
We modified the setting from 'Zero_histroy' mode to 'Always_on' mode in the COILOUTF filters for all of suspensions.
Also, we loaded the setting for all COILOUTF filters around 13:30 JST 4/25 2024, except for ETMX, ITMY, PRM, PR2 which someone use in this time.
According to the health check in January, gain of OSEM #3 (H2) and #4 (V3) are 3 dB and 5 dB smaller than before the earthquake, respectively (fig1 and 2).
Somehow, SRM IM OSEM TFs have lower gain (2-5 dB) than before the earthquake not only in H2 and V3 but also the others (fig3: case of H3).
Further investigation is necessary.
According to the health check in January, gain of OSEM #1 (H3) has almost no change before and after the earthquake.
According to the health check in January, gain of OSEM #2 (H1) is about 1dB smallerthan before the earthquake (fig1).
I performd health check of IMMT2 (fig1 - fig7).
All TFs seem fine.
I performd health check of IMMT1 (fig1 - fig7).
All TFs seem fine.
[Ikeda, Takahashi]
We checked the OSEMs visually. The flap of OSEM#3 and #4 in the IM was rotated more than 40°.
According to the health check in January, gain of OSEM #3 (H2) and #4 (V3) are 3 dB and 5 dB smaller than before the earthquake, respectively (fig1 and 2).
Somehow, SRM IM OSEM TFs have lower gain (2-5 dB) than before the earthquake not only in H2 and V3 but also the others (fig3: case of H3).
Further investigation is necessary.
[Ikeda, Takahashi]
We checked the OSEMs visually. The flap of OSEM#2 in the IM was rotated more than 40°.
According to the health check in January, gain of OSEM #2 (H1) is about 1dB smallerthan before the earthquake (fig1).
[Ikeda, Takahashi]
We checked the OSEMs visually. The flap of OSEM#1 in the IM was rotated more than 40°.
According to the health check in January, gain of OSEM #1 (H3) has almost no change before and after the earthquake.
[Takahashi, Ikeda, Hirata, Ushiba]
We checked the IMMT2 suspension visually. There was a small margin of the Pico-motor range for the pitch direction. There was no margin for yaw CCW direction (<1mm). There were not any rubbing magnets in the IM.
[Takahashi, Ikeda, Hirata, Ushiba]
We checked the IMMT1 suspension visually. There were margins of the pico-motor range much enough to adjust both pitch and yaw motion. There were not any rubbing magnets in the IM. When we took the pictures with a fiber scope touching the EQ stop frame, the TM pitch jumped due to a weak joint in the X-Y stage supporting the frame.
[Hirata, Ushiba-san, Ikeda-san]
We checked IR beam position around PR2 HR side Mid-size baffle. (How to make the vertical line is same as klog:29282)
It looks that IR beam positon is about 3.5mm +Y side away from the center of aperture.(IR beam center is 68.5mm and aperture center is 72mm on the ruler.)
The readout of the water fluid has recovered.
[YokozaWashimi, Tanaka, Ozaki, Sudo]
Today we tried to evaluate the seismic isolation of the OMC stacks, using a 3-axial accelerometer (S2315344) and an impact hammer (G1910656).
This is a quick report.
[Hirata, Ushiba]
We reconstructed POP forward beam path.
Though beam is still close to the knob of mirror mount just after periscope, the beam doesn't seem clipped.
Now, IR beam hits almost center of the both QPDs when the alignment to PR2 is good.
First, we requested aLIGNED state for IMMT1, IMMT2, and PRM.
Then, to confirm the good alignment has been kept from yesterday, we checked the beam spot on the PR2 HR target, which is almost center (fig1): good.
After confirming the alignment is good, we checked beam at high power beam dump (fig2).
Beam path seems far enough from the beam dump: also good.
After confirming the IR alignment to PR2, we checked beam spot on in-vac POM behind PR2 (fig3)
It is hard to say from the picture, but no clip happens at in-vac POM in my eyes.
After confirming beam spot on in-vac POM is not so bad, we started reconstruction of POP forward beam path.
At upper mirror on the periscope, the beam is not center but it seems no clip (fig4), so we keeep it as it is.
Then, we moved upper periscope actuators and hit the beam on the mirror center of lower mirror on the periscope (fig3).
After that, we moved FST1 in JGW-T1909623-v11, because it is hard to avoid clipping without moving the mirror.
Then, we aligned the beam to lower right of the FST1 (fig6) to avoid clipping at the knob of the mirror mount.
Figure 7 shows the current beam spot near the knob, which seems not so bad in my eyes.
After that, we realigned all the downstream and centered the QPDs.
During the work, on of the actuators of lower mirror mount on the periscope hits the periscope itself and cannot rotate to CW direction (we can rotate it in CCW direction).
So, we cannot move up the beam spot on FST1 without moving periscope mirror upward.
Since I am not familiar with the periscope installed for the POP forward, I didn't lift it up and keep it as it is.