Fabian, Yoshinori,
Log on March 11th, 2019,
= PRM =
-- As reported in the previous log, we tried the following*
> == Next ==
> -- we will compensate the yaw shift by the traverser so that the OpLev light is on the QPD.
> -- after these work, all the transfer functions will be measured.
-- By following the procedure reported in klog#581, we slided the traverser in +Yaw by 156 mdeg('= 30mdeg*6 - 10mdeg*3 +3mdeg*2) si that the oplev was centered.
-- The used script was the followinfg:
/kagra/Dropbox/Subsystems/VIS/TypeBp/Traversers/cont_serial.py
-- Then we measured some transfer functions of the IM-stage, and found some weird stuffs:
-- one wierd peak in a vertical transfer function (This looked a coupling from IM-L or T).
-- Actually, one OSE flag was close to the cavity though we cannot shift the OSEM plase anymore (There was no torelance anymore). Tilting BF-body from client machine could not solve this issue. The actual flag position should be checked visually.
-- the peak at 4.1 Hz was shifted to 4.4 Hz which was RM-P mode. and the frequency of the notch in damping filter was changed along with this, and the gain was slightly changed (from 1 to 0.8).
-- Now all the filters can be closed though, this system seems to include weird issues. Some more (visual) check would be necessary.