VIS (General)ryutaro.takahashi - 14:59 Thursday 24 March 2016 (1185)
Print this reportTest of 2.6m maraging rod (2)Load dependence was measured using a laser position sensor. Extension of 2.84mm was observed for the load of 272.8kg. Calculated expansion is 2.37mm (rod) + 0.14mm (neck) + 0.24mm (top plate) = 2.75mm. The amount of expansion is consistent with the calculation.