MIF (ITF Control)
takashi.uchiyama - 20:38 Monday 20 May 2019 (8909) Print this report
Comment to Alignment of X arm and Y arm cavities: Consistency between two arms (8818)
Toshikazu Suzuki

Suzuki-sama measured height of ITMX/Y by auto level.

ITMX is -0.9mm from the duct center,
ITMY is +0.5mm from the duct center,

Before height adjustment,
ITMX was +5.2mm from the duct center.
ITMY was -6.0mm from the duct center.

So that
ITMX was moved -6.1mm by the height adjustment works.
ITMY was moved +6.5mm by the height adjustment works.

ITMX is +5.8mm from the beam line measured by Telada-san.
ITMY is -2.0mm from the beam line measured by Telada-san.