keiko.kokeyama - 14:20 Tuesday 19 March 2019 (8449) Print this report
PRM misaliement [-8000, -9500] to hit the beam dump

[Ge, Kokeyama]

When you misalign PRM, the coil offset on IM should be
[-8000, -9500]
for the direct reflection to hit the high power beam dump. PSD is centered with this offset.



Today 12-1pm, we opened the IFI's viewport window and directly confirmed that the reflected beam hit the beam dump with a sensor card. The viewport was now closed.