Reports 1-1 of 1 Clear search Modify search
AOS (Baffles & Beam dumps)
tomotada.akutsu - 15:11 Wednesday 27 February 2019 (8223) Print this report
WAB at EYC finalization (cont'd)

K. Tanaka, Akutsu (continued from 8214)

  1. Did tiny remaining works for adapter block to the cooler head: torqued the four M8-40 screws with the plain washers and the spring washers. The final torque of 15Nm (5-> 10-> 15) was applied to each by a torque wrench.
  2. Did tiny remaining works for 6N Al heat links: torqued in 5Nm for all the screws (Fig 1).
  3. Did tiny remaining works for a thermometers: replaced the M3-8 screw with a M3-10 screw, which was used to fix a thermometer S/N#368, that was fixed at the WAB's suspension structure.

Then  connected electric wires of the thermometers: used P23 burndy connector, and the pin assign is (Fig 2):

    ABCD: V-, V+, A-, A+ in this order, and the pins are from the WAB itself
    EFGH: V-, V+, A-, A+ in this order, and the pins are from the suspension structure

and checked the diode's directions at the edge of the cable outside. Fig 3 shows the burndy cable from the P23 connector of the EYC chamber; I today attached the white circular tag to indicate it is coming from P23.

Note: the burndy cable from P23 of EYC is not sufficiently long to reach the cryocon; so we would need additional extension cable other than the one for changing connectors (adaptoring).