Aso, Miyo, Fujii with help of Sato,
= PR2 =
-- As repoted in klog#6504 and klog#7588, sometimes the signals of BF-LVDT-V2 and BF-LVDT-H3 were disappeared from the MEDM screen.
-- It was found that the climping press was not enough for those (hand-made) cables. Also, one of those cables had been injured close to Dsub9-connector. These had caused this issue.
-- Those cables were repaired by climping again with a lot of help of Sato-san.
-- After this cable repairing work, we confirmed that all the hand-made cable (see the following) connections looked good, i.e., this problem did not show up again. (this test was done by toching to the cable connectors by hand and electrical continuity was checked):
-- BF-LVDT-V1/V2/V3/H1/H2/H3
-- SF-GAS (*1)
(*1) Check for BF-GAS connector was done just by looking at it though.