Reports 1-1 of 1 Clear search Modify search
VIS (EX)
lucia.trozzo - 17:32 Wednesday 12 December 2018 (7348) Print this report
Drift of the vertical GAS Filter

[Lucia,Eleonora]

Today we noticed that respect to yesterday the signals F1_LVDT,BF_LVDT,F0_LVDT and BF_DAMP_V have drift (Pic1). This drift is due to the temperature variation in X end station because, yestarday (in the afternoon), the air conditioning system didn't work properly and it was switched off.

To compensate this drift and restore the old vertical position of the BF_DAMP_V we moved the fishing roads of F1 and F0 (see Pic2).

Images attached to this report