VIS (SRM)
fabian.arellano - 18:06 Monday 03 December 2018 (7269) Print this report
F0 LVDT condition

Today my aim was to lower the damper ring to its nominal position above to the SF and assess by how much the SF moves due to a possible magnetic force. The displacemt would be a change in the F0 LVDT readout.

Because I wanted to do the measurement in one go, I moved back the receptacles to their previous position, e.i. the bottom faces flushed with the bottom face of the F0 base plate.

Unfortunatelly, F0 LVDT is not providing a sensible output:

  1. A two point calibration yielded the factor 25 µm/cnt, whereas the factor previously determined was 0.147 µm/cnt (entry 7182).
  2. The polarity of the measurement was reversed: when the keystone goes up, the change in number of counts is negative.
  3. The resonant frequency is approximately 1.08 Hz (measured by visual inspection) instead of the 0.56 Hz measured before (entry 6886).
  4. The temperature today was 23.7 ºC.
  5. The notebook notes are here.

Apparently no wire was detached from the connector. In one week I'll check the health of the LVDT driver. I released the IP.

Comments to this report:
mark.barton - 17:07 Tuesday 04 December 2018 (7277) Print this report

Note that the 0.56 Hz frequency is for the SF and BF filters locked, whereas as of last Friday 11/30 they were both unlocked.