Reports 1-1 of 1 Clear search Modify search
IOO (General)
taiki.tanaka - 22:56 Wednesday 31 October 2018 (6798) Print this report
modulation index of f1 of EOM2a
Kohei Yamamoto, Taiki Tanaka


We measured modulation index of f1 of EOM2a with sweeping frequency of RF signal( see the plot).
-shut out delay line and b-path of 1st MZM
-take beat note with 1st refraction beam of AOM
In this measurement, the voltage on the EOM can be assumed like the following.
(Distribution amp) + (RF amp) - (Attenuator) - (Cable loss) = 13[V] + 17[V] - 7[V] - (Cable loss)

We will measure all the modulation indices tomorrow.
Images attached to this report
Search Help
×

Warning

×