With Enzo, Hatoya-san, and Hirata-san.
Today:
- Enzo and Hatoya-san had originally planned to experiment with different damper gaps by raising and lowering the top keystone, but found a curious coupling from keystone height to OL yaw.
- Rather than spend time resetting the OL, they decided to adjust the damper gap by raising the magnet ring.
- Hirata-san craned off the lid of the tank so they could access the top filter.
- They made a record of the current receptacle heights (see paper log book or photo thereof in SR3 Installation album of Google Photos) and then raised them 1 mm to give a wider gap.
- We left the tank lid off over night and set IP yaw TFs going.